SignalKine™ Sandwich ELISA Protocol
Specific for product: SignalKine™ Human IL-8 Sandwich ELISA Kit #7898.
This ELISA was developed for the detection and quantification of cytokines, chemokines and growth factors in cell culture media, serum and plasma. Recombinant protein standards are run alongside samples. Concentration of target in samples can be determined from the standard curve. We recommend running replicates of standards and samples.
Reagent Preparation
All reagents should be brought to room temperature before use.
- Wash Buffer:
- Prepare 1X wash buffer by diluting ELISA Wash Buffer (20X) in dH2O.
- SignalKine™ Assay Diluent A01:
- Buffered solution supplied as a 5X solution. 10 µl of Assay Diluent 1 is added to each well before addition of 40 µl of standard and samples.
- SignalKine™ Sample Diluent S01:
- Diluent for diluting standards and samples when working with plasma (citrate), plasma (heparin), or serum samples.
- SignalKine™ Sample Diluent S02:
- Diluent for diluting standards and samples when working with plasma (EDTA) samples.
- hIL-8 Detection Rabbit mAb (Biotinylated):
- Supplied lyophilized. Reconstitution with 60 µl of dH2O yields a 100X Concentrated Stock solution. Incubate at room temperature for 15 min with occasional gentle mixing to fully reconstitute. Dilute 100X solution 1:100 in Detection Antibody Diluent (green solution). A 96 well plate will need 50 µl/well of 1X solution, so approximately 5 ml of reagent is needed. Dilute 50 µl of 100X stock into 5 ml of Detection Antibody Diluent or dilute as much as needed if not using full 96 well plates.
- HRP-linked Streptavidin:
- Supplied lyophilized. Reconstitution with 60 µl of dH2O yields a 100X concentrated stock solution. Incubate at room temperature for 15 min with occasional gentle mixing to fully reconstitute. Dilute 100X solution 1:100 in HRP Diluent (red solution). A 96 well plate requires 50 µl/well of 1X solution, so approximately 5 ml of reagent is needed. Dilute 50 µl of 100X stock into 5 ml of HRP Diluent or dilute as much as needed if not using full 96 well plates.
- TMB Substrate:
- Supplied as a single component colorimetric HRP substrate. A blue color reaction is indicative of the presence of HRP and a positive result.
- STOP Solution:
- STOP solution is added to TMB substrate to stop the HRP reaction. STOP solution converts blue solution to yellow, which is then read on a microplate reader at 450 nm.
Human IL-8 Standard Preparation
Add 0.1 ml dH2O to reconstitute Human IL-8 Standard (Lyophilized) (10,000 pg) to yield a 100,000 pg/ml solution. Let sit at room temperature for 15 min with occasional gentle mixing to fully reconstitute standard protein.
For detection in plasma (citrate), plasma (heparin), and serum, samples and protein standards are diluted with SignalKine™ Sample Diluent S01, (provided in kit). For plasma (EDTA), samples and protein standards are diluted in SignalKine™ Sample Diluent S02, (provided in kit). For cell culture media, samples and protein standards are diluted in fresh cell culture media corresponding to sample media (not supplied).
A standard range of 1000, 500, 250, 125, 62.5, 31.25, 15.6 and 0 pg/ml is set up. This can be achieved by diluting the protein standard, (100,000 pg/ml) 1:100, which yields 1000 pg/ml. From the 1000 pg/ml solution, serial two-fold dilutions are performed to achieve the other standard concentrations.
We recommend diluting 10 µl of the 100,000 pg/ml protein standard into 990 µl of either SignalKine™ Sample Diluent S01, SignalKine™ Sample Diluent S02 or cell culture medium to yield the 1000 pg/ml standard. Serial two-fold dilution can be achieved by setting up six tubes with 200 µl of the appropriate diluent and then serially transferring 200 µl from the preceding tube. Mix well before each transfer. Use the appropriate diluent as the 0 pg/ml standard. (See graphic below).
Sample Preparation
This assay has been validated with cell culture medium, plasma (citrate), plasma (EDTA), plasma (heparin), and serum samples. After sample collection, store in single use aliquots at −20°C. Avoid multiple freeze/thawing.
Particulates in cell culture media should be removed by centrifugation. Note any contaminated, clotted, or hemolyzed samples and interpret these results judiciously.
Absorbance values falling within the hIL-8 standard curve range can be determined directly from the standard curve. Sample absorbance values higher than the 1000 pg/ml hIL-8 value should be diluted so that the absorbance values fall into standard curve range.
Procedure
- Once microwell strips have reached room temperature, take out as many strips as needed and insert into strip holder. Unused microwell strips should be resealed and stored at 4°C.
- Add 10 µl of 5X SignalKine™ Assay Diluent A01 to each well.
- Add 40 µl of hIL-8 Standards and samples to the wells. Cover with plate sealer and incubate for 2 hrs at room temperature.
- Wash plates 4 times with 1X Wash Buffer.
NOTE: Maximum wash volume is 170 µl/well. These are low volume microwells and have a smaller volume capacity than full size microwells.
- Add 50 µl of 1X hIL-8 Detection Rabbit mAb (Biotinylated) per well. Cover with plate sealer and incubate for 1 hr at room temperature.
- Wash plates 4 times with 1X Wash Buffer.
- Add 50 µl of 1X HRP-linked Streptavidin per well. Cover with plate sealer and incubate for 30 min at room temperature.
- Wash plates 4 times with 1X Wash Buffer.
- Add 50 µl of TMB substrate per well. Cover with plate sealer and incubate for 10 min at room temperature.
- After the 10 min incubation, add 50 µl of STOP solution per well.
- Read on a microplate reader, absorbance wavelength set at 450 nm. Plate should be read within 30 min.
Calculation of Results
Average replicates for standards and samples.
Using a graphing/curve fitting software, plot a 4-parameter logistic curve fit. A standard curve can also be generated by plotting the standard concentration on the x-axis and the absorbance corresponding to the standard concentration on the y-axis. A best-fit curve is drawn through the points on the graph.
Sample concentrations are determined by interpolating from the standard curves. Account for sample dilution by multiplying determined concentration by dilution factor.
Typical Data
Represented on right are hIL-8 Standard Curves, one diluted in SignalKine™ Sample Diluent S01, a second diluted in SignalKine™ Sample Diluent S02 and a third in cell culture media (RPMI + 10% FBS). Although these are typical of the standard curves that will be generated using this kit, a new set of standards should be run for each new experiment.