Basket Updated

0

Items added

1% for the planet logo
< Back to Support Article Search Results

Can I set my centrifuge to RPM rather than RCF for spinning my samples generated using the Cell Fractionation Kit #9038?

It is important to use the RCF (relative centrifugal force) settings on your centrifuge rather than the RPM (revolutions per minute) settings when using the Cell Fractionation Kit #9038 as RCF accounts for the size of the rotor and is therefore a more accurate representation of the force being applied to the sample. 

Last updated: February 26, 2024

Was this article helpful?

Technical Support

Email: [email protected]

Send Us a Message

Call: 877-678-8324

Customer Support

Email: [email protected]

Send Us a Message

Call: 877-616-2355

Fax: 866-432-6112

Contact Sales

Email: [email protected]

Send Us a Message

Call: 877-616-2355