Can the SARS-CoV-2 Spike Protein Serological IgG ELISA Kit #20154 be used for quantitation?

The SARS-CoV-2 Spike Protein Serological IgG ELISA Kit #20154 kit is meant to be used to determine if the sample is positive or negative for SARS-CoV-2 and not for relative quantitation. The included positive control is meant to be used as an on/off signal to confirm that the assay is working and should have an OD signal greater than 1.5. This is not meant to be used for standard curve generation or quantitation.
 