Render Target: SSR
Render Timestamp: 2024-12-15T03:30:44.700Z
Commit: 611277b6de3cd1bb065350b6ef8d63df412b7185
Cell Signaling Technology Logo
1% for the planet logo
< Back to Support Article Search Results

What is the best titration range to test in order to determine the optimal antibody dilution when using the ProteinSimple system?

When we validate an antibody for Simple Western™ we titrate the antibody dilution against a broad range of lysate concentrations. We recommend testing at least three antibody dilutions (1:10, 1:50 and 1:250) across a range of lysate concentrations (0.1 mg/mL – 1 mg/mL).

Last updated: September 12, 2024

Was this article helpful?