Render Target: SSR
Render Timestamp: 2024-12-15T20:32:26.197Z
Commit: 611277b6de3cd1bb065350b6ef8d63df412b7185
Cell Signaling Technology Logo
1% for the planet logo
< Back to Support Article Search Results

What is the expected signal strength for the PTMScan® Control peptides?

We have observed >1E6 peak area for each control peptide when used at the supplied concentration followed by PTMScan and PTMScan HS enrichment, as detected by our LCMS instruments.  However, the limit of detection (LOD) depends on many factors such as the sensitivity of the instrument and the LC gradient that is unique to each user.  Users may adjust the concentration of the supplied peptides as needed to find an appropriate signal strength that has a sufficiently high signal-to-noise ratio for reliable detection.
 

Last updated: September 12, 2024

Was this article helpful?

Technical Support

Email: [email protected]

Send Us a Message

Call: 877-678-8324

Customer Support

Email: [email protected]

Send Us a Message

Call: 877-616-2355

Fax: 866-432-6112

Contact Sales

Email: [email protected]

Send Us a Message

Call: 877-616-2355