Render Target: SSR
Render Timestamp: 2024-07-27T01:32:15.399Z
1% for the planet logo
< Back to Support Article Search Results

What mass tolerance should I use to detect the PTMScan® Control Peptides?

We recommend users to use the appropriate mass tolerance window based on historical performance of their LCMS instrument, the settings of the acquisition method and elapsed time from their most recent mass calibration.  For instance, when we acquire our data on a Q-Exactive with lock-mass activated, we generally use a tolerance window of 2.5ppm for the extracted ion chromatograms of the Control Peptides.  Generally, a window of up to 5ppm should be appropriate for most users operating an Orbitrap based instrument that is recently calibrated (i.e. within a week).  For Ion-Trap based instruments, users may need to use a higher mass tolerance window.
 

Last updated: February 26, 2024

Was this article helpful?

Technical Support

Email: [email protected]

Send Us a Message

Call: 877-678-8324

Customer Support

Email: [email protected]

Send Us a Message

Call: 877-616-2355

Fax: 866-432-6112

Contact Sales

Email: [email protected]

Send Us a Message

Call: 877-616-2355