Render Target: SSR
Render Timestamp: 2024-12-12T12:49:19.119Z
Commit: 611277b6de3cd1bb065350b6ef8d63df412b7185
Cell Signaling Technology Logo
1% for the planet logo
< Back to Support Article Search Results

What can cause false positives using the Senescence β-Galactosidase Staining Kit #9860?

The pH of the staining solution is critical and a pH below 6.0 is the most common cause of false positives when using the Senescence β-Galactosidase Staining Kit #9860. False positives may also be seen with:

  • Serum-starved cells
  • Contact inhibited cells
  • Confluent or over-confluent cells 
  • Increased lysosomes unrelated to senescence
  • Specific cell lines

Last updated: September 12, 2024

Was this article helpful?

Technical Support

Email: [email protected]

Send Us a Message

Call: 877-678-8324

Customer Support

Email: [email protected]

Send Us a Message

Call: 877-616-2355

Fax: 866-432-6112

Contact Sales

Email: [email protected]

Send Us a Message

Call: 877-616-2355