What can cause false positives using the Senescence β-Galactosidase Staining Kit #9860?

The pH of the staining solution is critical and a pH below 6.0 is the most common cause of false positives when using the Senescence β-Galactosidase Staining Kit #9860. False positives may also be seen with:

  • Serum-starved cells
  • Contact inhibited cells
  • Confluent or over-confluent cells 
  • Increased lysosomes unrelated to senescence
  • Specific cell lines

Last updated: September 12, 2024